ClearanceBrief
Recall Z-1130-2025 ยท OFX General Purpose Dental Tray

Cardinal Health 200, LLC

Cardinal Health Presource Kits: 1) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF104; 2) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF105 3) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF106 4) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF107 5) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF108 6) KIT,MAXILLOFACIAL,PRESBYTERIAN, Catalog Number:PWV2MF109 7) KIT,

DateFeb 10, 2025 (initiated Nov 8, 2024)
StatusOpen, Classified
ReasonPotential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.
Root causeMaterial/Component Contamination
ActionAn URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via
Quantity973,785 total units
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.