Recall Z-1172-2025 ยท DWJ System, Thermal Regulating
Augustine Temperature Management, LLC
Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses.
| Date | Feb 14, 2025 (initiated Feb 15, 2024) |
| Status | Open, Classified |
| Reason | Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact . |
| Root cause | Device Design |
| Action | Augustine Temperature Management contacted their consignees by visit beginning on 02/15/2024 to replace the affected cables. The firm collected email confirmation that the original cables had either been returned or destroyed. If you have any questions, contact the Product Manager at 952-465-3526 or email: |
| Quantity | 7 unit |
| Distribution | US distribution to states of: AL, TX, NV |
| FDA record | Open on fda.gov |