ClearanceBrief
Recall Z-0712-2025 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Maquet Cardiovascular, LLC

Vasoview Hemopro 2 (w/Vasoshield) Endoscopic Vessel Harvesting System. Model Numbers: VH-4001. Indicated for use in minimally invasive surgery allowing access for vessel harvesting.

DateJan 3, 2025 (initiated Dec 9, 2024)
StatusOpen, Classified
ReasonPotential for two issues: 1. Bent or detached heater wire; 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool
Root causeDevice Design
ActionGetinge notified consignees on 12/09/2024 via Urgent Medical Device Correction letter posted digitally on Getinge.com and sent to consignees via letter. Consignees were not instructed to return affected units, but instead to review the IFU information provided with the letter to minimize over-delivery of energy and mitigate risks. Getinge instructed consignees to notify customers if further distributed, as well as requested to complete and return the MEDICAL DEVICE CORRECTION RESPONSE FORM. Training to reinforce precautions and appropriate use of the device is expected to be deployed and co
Quantity11,101 units US
DistributionNationwide distribution. International distribution to Australia, Austria, Bahrain, Bangladesh, Belgium, Canada, Czech Republic, Denmark, Finland, France, Germany, Guatemala, Hong Kong, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Kuwait, Malaysia, Mexico, Netherlands, Netherlands, New
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101274VASOVIEW HEMOPRO 2 ENDOSCOPIC VESSEL HARVESTING SYSTEM
Maquet Cardiovascular, LLC, Jun 11, 2010