ClearanceBrief
Recall Z-0816-2025 ยท CAD Vaporizer, Anesthesia, Non-Heated

GE Medical Systems, LLC

Tec 820 SEV Anesthesia Vaporizer, Reference Number 1177-9820-SEV

DateJan 7, 2025 (initiated Nov 18, 2024)
StatusOpen, Classified
ReasonDue to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Root causeProcess control
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 11/18/24 was sent to customers. Actions to be taken by Customer/User You can continue to use the vaporizer in accordance with the instructions in the User Reference Manual (URM) and the actions described below: 1. Perform the Output Concentration Check described in Appendix 1 for every impacted vaporizer (see Affected Product Details in this letter). " If a vaporizer passes the Output Concentration Check, you can continue to use that vaporizer. GE HealthCare recommends use of anesthetic agent monitoring during vaporizer use. " If a
Quantity928 units
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172702Tec 820, Tec 850
Datex-Ohmeda, Inc., Jan 4, 2018