Recall Z-0814-2025 ยท CAD Vaporizer, Anesthesia, Non-Heated

GE Medical Systems, LLC

Tec 6 Plus Anesthesia Vaporizer, Reference Numbers 1107-9601-000, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9610-000, 1107-9612-000, 1107-9616-000, 1107-9617-000, 2066835-001, 2069767-001, 2078800-001, 2091008-001, 2091009-001, M1098876, M1145610, M1220023, M1220854, M1221495, and M1233992

DateJan 7, 2025 (initiated Nov 18, 2024)
StatusOpen, Classified
ReasonDue to a manufacturing issue, certain Tec 6 Plus and Tec 800 series vaporizers can deliver less anesthetic agent than the vaporizer dial setting.
Root causeProcess control
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 11/18/24 was sent to customers. Actions to be taken by Customer/User You can continue to use the vaporizer in accordance with the instructions in the User Reference Manual (URM) and the actions described below: 1. Perform the Output Concentration Check described in Appendix 1 for every impacted vaporizer (see Affected Product Details in this letter). " If a vaporizer passes the Output Concentration Check, you can continue to use that vaporizer. GE HealthCare recommends use of anesthetic agent monitoring during vaporizer use. " If a
Quantity1573 units
DistributionUS Nationwide. Global Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000275DATEX-OHMEDA TEC 6 PLUS ANESTHESIA VAPORIZER, DATEX-OHMEDA TEC 6 PLUS NAD VARIANT ANESTHESIA
Datex-Ohmeda, Inc., Apr 12, 2000