Recall Z-0779-2025 ยท OFF Central Venous Catheter Tray
MEDLINE INDUSTRIES, LP - Northfield
Medline custom medical procedure kits labeled as: 1) VP SHUNT CDS, REF CDS983470D; 2) 1/2X1/2 NEURO SPONGE PACK-LF, REF DYNJ0713177; 3) BURR HOLE SHUNT PACK, REF DYNJ56819D; 4) BURR HOLE SHUNT PACK, REF DYNJ56819F; 5) BURR HOLE SHUNT PACK, REF DYNJ56819G; 6) RO VP SHUNT PACK, REF PHS392834010D; 7) VP SHUNT, REF SYNJ10307A.
| Date | Dec 30, 2024 (initiated Nov 21, 2024) |
| Status | Open, Classified |
| Reason | Custom convenience kits are being recall because the contain non-sterile Codman Surgical Patties & Surgical Strips, which have the potential for endotoxin contamination. |
| Root cause | Nonconforming Material/Component |
| Action | Medline issued a MEDICAL DEVICE VOLUNTARY RECALL notice on 11/21/2024 to its consignees via US Mail and email. The notice explained the issue, potential risk to health, and requested that the affected product be quarantined. Once the response is received, Medline will provide labels to place on affected inventory, with instructions for the consignee to remove the affected component prior to using the kit. |
| Quantity | 1402 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of UAE, and Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.