Recall Z-0686-2025 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Philips North America Llc

Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216

DateDec 17, 2024 (initiated Nov 20, 2024)
StatusOpen, Classified
ReasonThe following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless cl
Root causeUnder Investigation by firm
ActionConsignees were notified on about 11/20/2024 via certified mail. Consignees were instructed to identify affected units, share the communication with clinical staff, and place the Urgent Medical Device Correction with the documentation of the Philips Smart-hopping 2.0 AP 1.4 GHz and associated devices. A Philips representative will contact customers to arrange a software upgrade to E.19.02 (5.0.19.02) or greater to correct the issues. Distributors were instructed to contact customers and provide the Medical Device Correction to each.
Quantity4,811 units
DistributionNationwide distribution