Recall Z-0704-2025 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

GE OEC Medical Systems, Inc

OEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended to provide fluoroscopic and digital spot images of the patient anatomy, interventional tools/devices, and contrast agents during diagnostic, interventional, and surgical procedures.

DateDec 18, 2024 (initiated Nov 21, 2024)
StatusOpen, Classified
ReasonDue to manufacturing issue of insufficient sealing of X-ray tubes and can result in a potential oil leak and loss of fluoroscopic imaging capability.
Root causeProcess control
ActionOn 11/21/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter informing customers that GE HealthCare has become aware that recently manufactured OEC Elite and OEC 3D mobile C-arms, as well as recently replaced X-ray tubes on certain OEC Elite, OEC 3D and OEC 9900/OEC 9800 mobile C-arms potentially have insufficient sealing that can result in an oil leak. This issue could result in loss of fluoroscopic imaging capability. Customers are instructed to: 1.Continue to use their OEC Elite, OEC 3D, OEC 9900 and/or OEC9800 mobile C-arm. 2.If oil is found leaking from the X-ray tube on th
Quantity100 systems
DistributionU.S.: AK, AZ, CA, CO, CT, DC, FL, GA, Guam, HI, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Australia, Brazil, Canada, Chile, Costa Rica, India, Italy, Philippines, Poland, Portugal, Sweden,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K170752OEC Elite
Oec Medical Systems, Inc., May 19, 2017
K171565OEC Elite
Ge Oec Medical Systems, Inc., Aug 10, 2017
K172550OEC Elite
Ge Oec Medical Systems, Inc., Nov 16, 2017
K192819OEC Elite
Ge Oec Medical Systems, Inc., Nov 8, 2019