ClearanceBrief
Recall Z-0555-2025 ยท ONG Powered Laser Surgical Instrument With MicrobeamFractional Output

WINGDERM ELECTRO-OPTICS LTD.

Non-ablative Fractional Laser Systems, Model: WFB-01

DateDec 5, 2024 (initiated Aug 23, 2024)
StatusOpen, Classified
ReasonModel WFB-01 sold to US customers before Aug 23, 2022 has been found to have a software bug (6 units). All WFB-01 models sold to US customers found to have labeling non-compliances (22 units)
Root causeRadiation Control for Health and Safety Act
ActionThe firm sent a notification letter to the relevant distribution on June 6, 2024. Products sold before August 23, 2022, will require a software version upgrade to F1-B-1.1.012 or Higher.
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192350Medical Non-Ablative Fractional Laser Systems
Wingderm Electro-Optics , Ltd., Apr 30, 2020