Recall Z-0677-2025 ยท NXO Calprotectin, Fecal

Diasorin Inc.

DiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens

DateDec 12, 2024 (initiated Nov 7, 2024)
StatusOpen, Classified
ReasonCustomer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.
Root causeProcess control
ActionDiaSorn issued an Urgent Field Safety Notice - Medical Device Recall notice to its consignees on 11/07/2024 via email. The notice explained the issue with the product, risk to health, and requested the following: "Ac xons to be taken by the Customer/User " No need to review past patient results. " Devices should be inspected for loose clear caps prior to use. The user should confirm that the clear cap is tight by grasping the cap and twisting. If the cap is loose, the device should be discarded. Do not attempt to use the device as the loose clear cap may have allowed buffer to leak, causin
Quantity21905 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K213858LIAISON Calprotectin, LIAISON Q.S.E.T. Device Plus
DiaSorin, Inc., Jul 26, 2022