Recall Z-0795-2025 ยท CAW Generator, Oxygen, Portable

JIANGSU JUMAO X-CARE MEDICAL EQUIPMENT CO LTD

JMC5A Ni/TruAire-5 Oxygen Concentrator, Model: O2C5L

DateJan 3, 2025 (initiated Dec 9, 2024)
StatusOpen, Classified
ReasonOxygen concentrator may melt or catch on fire during use.
Root causeComponent change control
ActionOn 11/26/2024, a recall notice was sent a distributor asking them to take the steps below. On 12/09/2024, recall notices were sent from the distributor to customers who were asked to do the following: This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. All units from the specified serial number range MUST be immediately discontinued from use and removed from service and promptly returned. Complete and return the acknowledgement and receipt form via email to recall@c
Quantity6,520
DistributionUS: OH
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K230969Oxygen Concentrator
Jiangsu Jumao X-Care Medical Equipment Co., Ltd., Nov 9, 2023