ClearanceBrief
Recall Z-0726-2025 ยท OJH Orthopedic Tray

MEDLINE INDUSTRIES, LP - Northfield

Medline brand, medical procedure convenience kits, labeled as: 1) LAMINECTOMY CDS-LF, SKU CDS780045N; 2) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119U; 3) NEURO CRANIOTOMY CDS #36-RF, SKU CDS780119V; 4) CRANIOTOMY CDS, SKU CDS981753K; 5) LAMINECTOMY CDS-LF, SKU CDS981923X; 6) CRANIOTOMY, SKU CDS982719Q; 7) CRANIOTOMY, SKU CDS982719R; 8) CRANIOTOMY, SKU CDS982719S; 9) CRANIOTOMY, SKU CDS9827

DateDec 20, 2024 (initiated Nov 6, 2024)
StatusOpen, Classified
ReasonSurgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods
Root causeNonconforming Material/Component
ActionAn IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 11/7/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submi
Quantity22,970 kits
DistributionUS Nationwide. Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.