Recall Z-0725-2025 ยท OHD Gynecological Laparoscopic Kit
MEDLINE INDUSTRIES, LP - Northfield
KIT,SPINE LH Convenience kit, SKU DYNJ906448A.
| Date | Dec 20, 2024 (initiated Nov 6, 2024) |
| Status | Open, Classified |
| Reason | Surgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods |
| Root cause | Nonconforming Material/Component |
| Action | An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 11/7/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submi |
| Quantity | 160 kits |
| Distribution | US Nationwide. Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.