Recall Z-0720-2025 ยท OGR Ear, Nose, And Throat Surgical Tray

MEDLINE INDUSTRIES, LP - Northfield

Medline brand, medical procedure convenience kits, labeled as: 1) SINUS CDS-1, SKU CDS982901F; 2) NASAL PACK, SKU DYNJ17493G; 3) FESS PACK-POB, SKU DYNJ33520A; 4) SINUS PACK, SKU DYNJ44845G; 5) SINUS PACK, SKU DYNJ48330B; 6) ENT PACK, SKU DYNJ50572; 7) ENT PACK, SKU DYNJ53573A; 8) ENT HUTCH, SKU DYNJ58856A; 9) PACK NASAL, SKU DYNJ60111A; 10) PACK NASAL, SKU DYNJ60111B; 11) ENT CUSTOM PACK, SKU DYN

DateDec 20, 2024 (initiated Nov 6, 2024)
StatusOpen, Classified
ReasonSurgical Patties & Surgical Strips in specific lots of Custom Convenience Kits were found to contain higher-than-expected levels of endotoxin within the raw material used to produce sterile Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in those finished goods
Root causeNonconforming Material/Component
ActionAn IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 11/7/24 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Upon receipt of your submi
Quantity6467 kits
DistributionUS Nationwide. Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.