Recall Z-0685-2025 ยท MWJ Electrocardiograph, Ambulatory (Without Analysis)

Spacelabs Healthcare, Ltd.

Eclipse Mini REF 98900 The Eclipse MINI Model 98900 is a portable non-invasive continuous ambulatory ECG patch recorder intended to record the patient s electrocardiogram.

DateDec 17, 2024 (initiated Nov 25, 2024)
StatusOpen, Classified
ReasonDue to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
Root causeProcess control
ActionOn 11/25/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter to customers inform them that the Eclipse Pro and Eclipse MINI cardiac Holter monitors may have mismatched serial numbers between the label and what is programmed in the devices. This is due to a manufacturing error. Customers are instructed to: complete the following steps for each Eclipse device under your control. Following these steps will allow you to check if your device is affected. 1. Visit https://eclipseutils.cardiology.spacelabs.com/ 2. The first time you use this site you will need to register for an accoun
Quantity1,779 devices
DistributionU.S.: AR, CA, CO, DE, FL, FA, LA, MA, MI, MN, MT, NC, NE, NU OH, OR, PA, Puerto Rico, TN, TX, UT, VA, WA, and WI. O.U.S.: Argentina Denmark Kuwait Serbia Australia Dominican Republic Lebanon Singapore Austria Ecuador Lithuania Spain Belgium El Salvador Martinique Sweden Benin France Mexico Switzerl
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212317Eclipse MINI Model 98900
Spacelabs Healthcare, Ltd., Nov 7, 2022