Recall Z-0669-2025 ยท FZX Guide, Surgical, Instrument

Hologic, Inc

Hologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101

DateDec 11, 2024 (initiated Oct 10, 2022)
StatusCompleted
ReasonIFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
Root causeLabeling Change Control
ActionHologic issued Letter-Revised Sterilization Instructions for Sizer Set on 10/10/22 via Fed'X. The letter states reason for recall, health risk and action to take: Effective immediately, please steam autoclave F0101 in accordance with such IFU, subject to the following parameters. Any additional parameters contained in Hologic s IFU for F0101 should no longer be utilized. link to their Instructions for Use (IFU) below: https://summitmedicalusa.com/wordpress/wp-content/uploads/2021/05/40192-01_RevZ_IS-IFU_Multilingual.pdf Please also note that when sterilizing F0101, one (1) layer of wrapping
Quantity178 units
DistributionUS Nationwide distribution.