ClearanceBrief
Recall Z-0751-2025 ยท NKB Thoracolumbosacral Pedicle Screw System

SEASPINE ORTHOPEDICS CORPORATION

Brand Name: Daytona Small Stature Product Name: CoCr LockingCap 4.5mm, Small Stature, T25, BT Model/Catalog Number: 79-0002 Software Version: N/A Product Description: Locking Cap Component: No

DateDec 23, 2024 (initiated Oct 29, 2024)
StatusOpen, Classified
ReasonDue to manufacturing defects observed on Locking Caps that affect the ability for components to mate.
Root causeNonconforming Material/Component
ActionUpdate 12/18/2024 - The firm sent an updated Customer Notification Letter on 12/12/2024 to customers that have not responded to the initial notification on 10/29/2024. On 10/29/2024, firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter to customer informing them that the firm has identified a nonconformance with the threads of the locking cap components that may be incongruous with optimal manufacturing specifications. The nonconformance can result in a locking cap that does not mate appropriately with the screw tulip. Customers are instructed to: -Review their inventory and return the affec
Quantity92 devices
DistributionU.S. Nationwide distribution in the states of CA, CO, FL, GA, PA, TX, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K232566NewPort Spinal System; Daytona Small Stature Spinal System; Malibu Spinal System; Mariner Pedicle Screw Systems: Mariner
SeaSpine Orthopedics Corporation, Apr 19, 2024