Recall Z-0667-2025 · BSZ Gas-Machine, Anesthesia

Draeger Medical, Inc.

Brand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.

DateDec 18, 2024 (initiated Oct 15, 2024)
StatusOpen, Classified
ReasonThe Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Root causeProcess control
ActionOn October 15, 2024 Urgent Medical Device Correction letters were sent to customers. Necessary Action: The corrective action is to replace the ventilator motor assembly. You will be contacted by your local Draeger Service Representative to arrange a date for the ventilator motor assembly replacement. Until the corrective action is implemented, you may continue operating your devices with the customary care by only using the device under permanent supervision in line with the Atlan Instructions for Use. If the mechanical ventilation fails, use Man/Spont ventilation mode to ventilate the patient
Quantity280 units (OUS only)
DistributionNo Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K230931Atlan
Drägerwerk AG Co. Kgaa, Jul 23, 2023