Recall Z-0802-2025 ยท FMI Needle, Hypodermic, Single Lumen

Greiner Bio-One GmbH

VACUETTE SAFELINK, REF: 450210

DateJan 3, 2025 (initiated Nov 19, 2024)
StatusOpen, Classified
ReasonThe VACUETTE SAFELINK Holder with male luer lock may leak blood due to the protective rubber sleeve bunching up and not retracting when using the device.
Root causeProcess control
ActionOn 11/19/2024, customers and distributors were mailed and emailed recall notices asking them to do the following: 1) Stop using and isolate affected devices. 2) If the product has been further distributed or transferred to another location, notify those locations of the recall, and advise the name/location in the provided space on the attached Product Disposition Site Confirmation. 3) Complete and return the Product Disposition Site Confirmation form via email to recalls.us@gbo.com If you have additional questions, contact the firm's Technical Service Department at 800-515-8112.
Quantity188,800 pcs
DistributionUS Nationwide distribution in the states of IL, PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K180545VACUETTE SAFELINK Holder with male luer lock
Greiner Bio-One Na, Inc., May 21, 2018