ClearanceBrief
Recall Z-0519-2025 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Medical Systems, LLC

SIGNA Architect AIR, Nuclear Magnetic Resonance Imaging System

DateNov 22, 2024 (initiated Oct 28, 2024)
StatusOpen, Classified
ReasonGradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
Root causeDevice Design
ActionGE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees on 10/28/20224 delivered through a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken for the continued safe usage of the affected devices: "Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safet
Quantity3 units
DistributionWorldwide Distribution: US (nationwide) and OUS (foreign) to countries of: Canada, China, India, Japan, Korea, Latvia, Mexico, Poland, Russia, Saudi Arabia, Spain, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160618Discovery MR750 3.0T, Discovery MR450 1.5T, Discovery MR750w 3.0T, Optima MR450w 1.5T
Ge Medical Systems, LLC, Jun 9, 2016
K163331Discovery MR750 3.0T; Discovery MR750w 3.0T;Discovery MR450 1.5T; Discovery MR450w 1.5T; SIGNA Architect and SIGNA Artis
Ge Medical Systems, LLC, Mar 17, 2017
K202966SIGNA Architect
Ge Healthcare (Ge Medical Systems, LLC), Nov 13, 2020