ClearanceBrief
Recall Z-0661-2025 ยท GWM Device, Monitoring, Intracranial Pressure

Medtronic Neurosurgery

Becker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKER W/1-WAY VALVE 26040 KIT 26040 BECKER EDMS II 27609 BECKER 27609 W/NEEDLESLESS INJ. SITE 27670 BECKER 27670 STOP BELOW DC NDL-LS IN 27672 BECKER 27672 ONEWAY VLV Y-SITE PT LN 27702 EDMS 27702 EURO BECKER II GR PT LN 27761 BECKER 27761 STOP BCNDLESS PATLN 24 27767 BECKER 27767 NO STPCOCK ON P

DateDec 19, 2024 (initiated Nov 1, 2024)
StatusOpen, Classified
ReasonDue to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
Root causeUnder Investigation by firm
ActionOn 11/01/2024, the firm sent an URGENT: MEDICAL DEVICE CORRECTION Letter via US mail to customers informing them that through customer complaints, the firm is aware of potential cracking and leaking in the "stopcock" components on the Becker and Exacta External Drainage and Monitoring Systems. Customer are instructed to: Recommended Mitigations: -Prior to use, inspect all stopcocks and connections to ensure that connections are secure and there are no visible cracks in the stopcocks. If cracks or leaks are identified, do not use return the device to Medtronic. -The system must be pre-filled w
Quantity65,810 units
DistributionWorldwide distribution. US nationwide including Puerto Rico; Algeria, Argentina, Australia, Austria, Bahrain, Barbados, Bosnia And Herzegovina, Brunei Darussalam, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, Fren
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200456Becker External Drainage and Monitoring System
Medtronic, Inc., Mar 26, 2020