Recall Z-0674-2025 ยท FYC Gown, Isolation, Surgical

Cardinal Health 200, LLC

Convertor Isolation Gown, AAMI Level 3 protection, Product Codes AT6100, AT6100-XL, AT6200, and AT6200-XL

DateDec 11, 2024 (initiated Nov 1, 2024)
StatusOpen, Classified
ReasonIsolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (The Association for the Advancement of Medical Instrumentation) Level 3 protection requirements.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE PRODUCT RECALL notification letter dated 11/1/24 was sent to customer. Actions Required: 1. REVIEW your inventory for the affected product codes. Location of product code and is shown on the label below (Attachment 1). 2. COMMUNICATE with all personnel that utilize the above listed Cardinal Health AAMI Level 3 Isolation Gowns. 3. SEGREGATE and quarantine all affected product upon review of your inventory. Affected product should not be used. Utilize return directions below to return product. 4. DISSEMINATE this notice to all departments, clinics and external campuses
Quantity940,203 total units
DistributionUS Nationwide. Canada, UAE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160339Cardinal Health Isolation Gown
Cardinal Health200, LLC, Jul 28, 2016