Recall Z-0579-2025 ยท OWB Interventional Fluoroscopic X-Ray System

Orthoscan, Inc.

Ziehm Vision RFD 3D. Interventinoal fluoroscopic x-ray system

DateDec 2, 2024 (initiated Aug 8, 2024)
StatusOpen, Classified
ReasonZiehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit.
Root causeRadiation Control for Health and Safety Act
ActionZiehm Imaging, Inc. intends to notify the affected customers directly in the following manner: 1) Letter with return receipt requested, envelope, return, and reply to questionnaire, or E-mail contact. 2) Direct phone calls. Once contact is made and confirmed Ziehm Imaging Inc, shall provide replacement hand switch and instructions to the end user. Wired Hand switch 88036 Vl.02 will be requested to be returned by the customer/end user to Ziehm Imaging Inc. and will be subjected to be scrapped.
Quantity50
DistributionU.S. (including Puerto Rico).
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K231701Ziehm Vision RFD 3D
Ziehm Imaging GmbH, Jul 10, 2023