Recall Z-0693-2025 ยท EOQ Bronchoscope (Flexible Or Rigid)

Noah Medical

Galaxy System (GAL-001)

DateDec 18, 2024 (initiated Oct 25, 2024)
StatusOpen, Classified
ReasonBronchoscopic system uninterruptable power may have defective component, may cause power supply short circuit/trip protective circuit breakers causing sudden, unrecoverable power loss, which could occur while system is on mains or internal battery power; if during procedure, could cause interrupted/aborted procedure, may necessitate manual removal, which could lead to lung injury/pneumothorax.
Root causeDevice Design
ActionOn 10/25/24, the firm started hand delivering correction notices to customers informing them of the following: If your device is affected the firm will contact you to inspect and if necessary to replace the power supply. Firm contact: customerservice@noahmed.com, 1-888-325-6624
Quantity26
DistributionUS: CA, MA, PA, NJ, ID, CT, MO, NC, MD, IL, TN, VA, OH, IN, WV, NY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223144Galaxy System
Noah Medical Corp., Mar 1, 2023