Recall Z-0730-2025 ยท MAV Syringe, Balloon Inflation

Abbott Vascular Inc

20/30 INDEFLATOR, REF: 1000184, for use during cardiovascular procedures in conjunction with interventional devices such as balloon dilatation catheters to create and monitor pressure in the balloon.

DateDec 20, 2024 (initiated Nov 6, 2024)
StatusOpen, Classified
ReasonInflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.
Root causeComponent change control
ActionOn 11/6/2024, recall notices were distributed to customers asking them to take the following actions: 1) Stop using affected devices. 2) Complete and return the acknowledgement form. 3) Return unused devices to the firm. 4) Share this notification with relevant personnel in your organization. 5) If you have further distributed/transferred the affected products, notify those customers. If you have any questions contact firm customer service at 800-227-9902
Quantity1,418
DistributionUS: IA, MD, DC, IN, CA, NV, TX, CT, GA, MS, NY, OH, FL, NC, MI, KY, PA, VA, IL, OK, TN, UT, WA, WI, MT, OR, LA, HI, NE, AR, MO, MA, KS, ID, AL, AZ, PR, CO, SC. OUS: IN, KR, SG, TH, VN, AR, BR, CA, CO, PA, UY, BE, CZ, DE, ES, FR, GB, IT, PT, RE, SE, SK, TM, JP.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961471INDEFLATOR 20/30 INFLATION DEVICE AND 20/30 PRIORITY PACK (WITH THE .096 OR .115 ROTATING HEMOSTATIC VALVE)
Advanced Cardiovascular Systems, Inc., May 8, 1996