Recall Z-0987-2025 ยท MOM System, Hemodynamic, Implantable

St. Jude Medical

Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Model MN6000) (Which interfaces with CardioMEMS" Hospital System model CM3100) The CardioMEMS" HF System provides pulmonary artery (PA) hemodynamic data used for monitoring and management of heart failure (HF) patients. The system measures changes in pulmonary artery (PA) pressure which physicians use to initiate or modify heart

DateJan 17, 2025 (initiated Nov 6, 2024)
StatusOpen, Classified
ReasonDue to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.
Root causeUnder Investigation by firm
ActionOn 11/06/2024, the firm sent an "URGENT: MEDICAL DEVICES CORRECTION", to customers to inform them that on September 30, 2024, the Merlin.net PCN HF Portal (MN6000) was migrated to the cloud database and following the migration, some customers reported that their Hospital Systems were unable to transmit to the Merlin.net PCN. User Action Requested: Abbott requests that clinicians take the following actions: The Impacted CM3100 Hospital System devices at your facility is/are listed in the enclosed Appendix. o For a Hospital System marked SESSIONS SENT no further correction is needed. o Hospital
Quantity25 systems
DistributionU.S.: CA, CT, GA, IL, MI, MN, NC, NJ, NY, OR, PA, TX, and WA. O.U.S.: N/A
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.