Recall Z-0678-2025 ยท KPS System, Tomography, Computed, Emission

Hermes Medical Solutions AB

Hybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0

DateDec 12, 2024 (initiated Oct 31, 2024)
StatusOpen, Classified
ReasonDue a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study.
Root causeDevice Design
ActionOn 10/31/2024, the firm emailed a "Field Safety Notice (FSN)" letter to customers informing them that when reconstructing a SPECT/CT study, in some cases, the SPECT and CT series are not correctly aligned. This may be due to separately performed acquisitions, or due to the camera operator re-setting the reference point (landmark) between SPECT and CT acquisitions. Normally, the Hybrid Recon application warns the user if the Frames of Reference do not match, however in some configurations this warning message may be turned off. Customers are instructed how to avoid the problem by: In all cases
Quantity778 systems
DistributionU.S.: CA, CT, FL, HI, KY, MA, MD, MO, NJ. NY, OH, TN, TX, VA, and WA O.U.S.: Germany, and United Kingdoms
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171681Hermes Medical Imaging Suite v5.7
Hermes Medical Solutions AB, Nov 21, 2017