Recall Z-0522-2025 ยท MVU Reagents, Specific, Analyte

Kreatech Biotechnology Bv

Kreatech PapV-16, Flu labeled ISH probe, to detect presence of HPV-16 DNA inserts in human tissue, Model Number 40V016V495

DateNov 22, 2024 (initiated Nov 1, 2024)
StatusOpen, Classified
ReasonThe DNA fragments for this specific lot number are too large, resulting in potential background around the cells due to suboptimal hybridization. An absence of signal could also be observed as a result of the suboptimal hybridization of the larger DNA fragments. The user may interpret potential background as a false positive result and an absence of signal as a false negative result, if proper positive and negative controls are not included.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 10/31/24 was sent to customers. Actions to Be Taken by the Customer/User 1. Do not use or continue to use the batch listed above. You will receive a replacement product as soon as possible. 2. Please dispose of any unused products according to your organization s disposal policies. 3. Review and follow up on any results obtained using the above-mentioned product batch. 4. Please read and retain this communication. Please confirm receipt of this letter as soon as possible by signing and returning the below Acknowledgement Form to: QA.A
Quantity23 units
DistributionUS States: FL, NY
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.