Recall Z-0620-2025 ยท FDS Gastroscope And Accessories, Flexible/Rigid

Olympus Corporation of the Americas

EVIS EXERA III Gastrointestinal Videoscope GIF-1TH190. Model Number: GIF-1TH190

DateDec 5, 2024 (initiated Oct 31, 2024)
StatusOpen, Classified
ReasonHigh-level disinfection ("HLD") with Acecide-C disinfectant in the Olympus OER-Pro Automated Endoscope Reprocessor ("AER") has been removed as a compatible reprocessing method from the Instructions for Use when reprocessing the GIF-1THI90 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE ( GIF-1TH190 ).
Root causeUnder Investigation by firm
ActionOlympus notified consignees via letter on about 10/31/2024. Consignees were informed of the issue and of the upcoming revision of the labeling such that the GIF-1TH190 is no longer a compatible endoscope for reprocessing with Acecide-C in the OER-Pro. They were instructed that they would be contacted by their Olympus sales representative within the month to discuss potential OER-Elite and GIF-1TH190 transition programs. Additionally, they were instructed to carefully read the notification, ensure reprocessing staff and other users of the OER-Pro AER are aware of the corrective action, ackno
Quantity1,770 units
DistributionUS Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K232997EVIS Exera III Gastrointestinal Videoscope GIF-1TH190
Olympus Medical Systems Corporation, Jun 19, 2024