Recall Z-0770-2025 ยท MCW Catheter, Peripheral, Atherectomy

Angiodynamics, Inc.

AURYON LASER SYSTEM 100-120 VAC. Model Number: EXM001. Product Number: EXM-2001-1100

DateDec 27, 2024 (initiated Nov 25, 2024)
StatusOpen, Classified
ReasonPotential for procedural delays or interruptions during use of the Auryon Atherectomy System due to the Auryon Atherectomy System not advancing to the Activation (Ready) mode.
Root causeUnder Investigation by firm
ActionAngioDynamics notified consignees on about 11/25/2024 via URGENT VOLUNTARY MEDICAL DEVICE CORRECTION letter. Consignees were instructed to located affected devices on hand, place the revised Operator's Manual with affected systems, remove and destroy the Operator s Manual that is presently with the Auryon Laser System, and notify any customers of the recall if the Auryon Laser System has been redistributed to another facility. Additionally, consignees were asked to complete and return the Reply Verification Tracking Form.
Quantity138 units
DistributionUS Nationwide distribution to the states of AR, AZ, CA, CO, CT, FL, GA, IL, IN, KY, LA, MD, MI, MN, MO, MS, NC, NJ, NM, NY, OH, PA, RI, TN, TX, VA, WA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K241553Auryon Atherectomy Catheter 1.7mm; Auryon Atherectomy Catheter 1.7mm, Hydrophilic Coating
Eximo Medical, Ltd., Jun 27, 2024