ClearanceBrief
Recall Z-0495-2025 ยท NPT Aortic Valve, Prosthesis, Percutaneously Delivered

Abbott Medical

Abbott Navitor Transcatheter Aortic Heart Valve, 27 mm, NVTR-27, Sterile

DateNov 20, 2024 (initiated Oct 17, 2024)
StatusOpen, Classified
ReasonA manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
Root causeUnder Investigation by firm
ActionA Medical Device Recall letter was hand delivered Oct. 21, 2024 to the one US consignee for 1 Impacted Valve. Nine foreign consignees were also notified for 26 Impacted Valves distributed outside of the United States (OUS). Customers are advised to return any remaining unused Impacted Valves, if applicable, and complete and return the accompanying Acknowledgement Form to Abbott. Abbott is executing further testing to verify that long-term durability of the Impacted Valves has not been affected and will communicate any changes to the recommendations set forth in this letter if needed by Q4 2025
Quantity5 OUS
DistributionNew York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.