Recall Z-0640-2025 ยท OJG Neurological Tray
MEDLINE INDUSTRIES, LP - Northfield
MEDLINE convenience kits labeled as: 1) BASIC NEURO PACK, REF CDS780202P; 2) BASIC NEURO PACK, REF CDS780202Q; 3) BASIC NEURO PACK, REF CDS780202R; 4) ACDF CDS, REF CDS941061M; 5) NEURO KIT, REF CDS981327Q; 6) ZALE V-P SHUNT CDS, REF CDS983186J; 7) NEURO PACK-LF, REF DYNJ0038440C; 8) NEURO PACK-LF, REF DYNJ0406347M; 9) NEURO-LF, REF DYNJ0484186M; 10) NEURO-LF, REF DYNJ0484186N; 11) NEU
| Date | Dec 5, 2024 (initiated Oct 9, 2024) |
| Status | Open, Classified |
| Reason | Medline is recalling medical convenience manufactured using certain lots of SafeAir Smoke Evacuation Pencil which were subsequently recalled by Stryker. |
| Root cause | Under Investigation by firm |
| Action | Medline issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 10/11/2024 via mail and email. The notice explained the problem with the Stryker component, potential risk, requested the kits be quarantined pending additional labeling and o respond through their portal: https://recalls.medline.com. Once the response is received, Medline will provide additional labeling to be applied to the affected kits which states the following: "Attention: R-24-219, This kit contains a recalled item, STRYKER SafeAir Smoke Evacuation Pencil, VPN: 0703-047-000, 0703-047-001, 0703-047-002, 0703-0 |
| Quantity | 12847 units |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.