Recall Z-0665-2025 ยท IZL System, X-Ray, Mobile

Micro-X Ltd.

MICRO-X Rover Mobile X-ray System, # MXU-RV35; MICRO-X Rover Mobile X-ray System, # MXU-RV71

DateDec 12, 2024 (initiated Nov 28, 2023)
StatusOpen, Classified
ReasonMobile x-ray system can report and deliver a lower post-exposure mAs value than what set by the operator.
Root causeDevice Design
ActionOn 11/28/23 Modification 17 was sent to customers telling them that Control Boards need to be replaced with Control Board PCBA, 12425-04 that will eliminate the issue. For technical assistance contact: support-us@micro-x.com, 1-206-249-8764 Firm asked distributors to do the following: On next visit please confirm that the bulletins were completed and return the documentation.
Quantity18
DistributionUS Nationwide Distribution: NJ, CA, PR
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211423Rover
Micro-X Limited, May 21, 2021