ClearanceBrief
Recall Z-0486-2025 ยท MWS Stabilizer, Heart

Maquet Cardiovascular, LLC

Acrobat-i Vacuum Stabilizer System. Model Number C-OM-10000Z

DateNov 18, 2024 (initiated Oct 16, 2024)
StatusOpen, Classified
ReasonDeviations in the manufacturing process created a compromise in the sterile barrier, in which sterility cannot be assured during the duration of the product's 2-year shelf life.
Root causeProcess change control
ActionMAQUET Cardiovascular LLC / Getinge notified consignees via letter FedEx 2-Day Delivery on 10/16/2024. Customers were instructed to forward the notification to all users and notify customers if further distributed, examine inventory and return any affected inventory, and to complete and return the MEDICAL DEVICE REMOVAL - RESPONSE FORM.
Quantity33,823 units
DistributionNationwide distribution. International distribution to Albania, Algeria, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brunei Canada, Croatia, Czech Republic, Denmark, Finland, France and Belarus, Germany, Greece, Hong Kong, India, Indonesia, Italy, Japan, Lebanon, Luxembo
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.