Recall Z-0489-2025 ยท LPS Urinary Homocystine (Nonquantitative) Test System

Roche Diagnostics Operations, Inc.

cobas HCYS, Homocysteine Enzymatic Assay, Material Number REF 06542921190, for use on cobas c 701/702, in vitro test for total L-homocysteine

DateNov 19, 2024 (initiated Oct 17, 2024)
StatusOpen, Classified
ReasonAbnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)
Root causeUnder Investigation by firm
ActionRoche issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 10/17/2024 via UPS Ground. The notice explained the issue, clinical significance, and requested the following actions be taken: "" HCYS kits lot number 747029 expired on September 30, 2024. Ensure any inventory you have of this lot has been discarded per your local guidelines. " Follow your laboratory s standard operating procedures to determine specific clinical implications for your patients and whether retrospective review of patient results is required. " Consult with the physician or pathologist at your faci
Quantity458 units
DistributionUS: NJ, TX, NY, FL, IN, TX, OK, SC, IA,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113793HOMOCYSTEINE TEST
Roche, Jun 5, 2012