Recall Z-0528-2025 ยท BSR Stylet, Tracheal Tube

Smiths Medical ASD, Inc.

smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP

DateNov 25, 2024 (initiated Oct 28, 2024)
StatusOpen, Classified
ReasonSmiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.
Root causeUnder Investigation by firm
ActionSmiths Medical issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 10/28/2024, mailed using a traceable means. The notice explained the issue, potential risk, and requested the consignee locate and destroy all affected products, and distribute the notice to all affected parties. Distributors were directed to notify their customers. For questions: About the communication: smithsmedical7602@sedgwick.com, 1-800-913-3897; To report adverse events: globalcomplaints@icumed.com, 1-866-216-8806
Quantity10,223 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.