Recall Z-0664-2025 ยท IZL System, X-Ray, Mobile

Micro-X Ltd.

MICRO-X Rover Mobile X-ray System, # MXU-RV35

DateDec 12, 2024 (initiated May 31, 2023)
StatusOpen, Classified
ReasonShots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.
Root causeSoftware design
ActionThe firm will notify consignees using a service bulletin that was already posted. Additionally, the firm will implement a firmware update to extend the timer (4 ms to 10 ms minimum) to ensure even short exposures achieve the necessary image quality for diagnostic use.
Quantity5
DistributionWorldwide and US Nationwide Distribution: NJ, PR, CA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211423Rover
Micro-X Limited, May 21, 2021