Recall Z-0988-2025 ยท MOM System, Hemodynamic, Implantable

St. Jude Medical

CARDIOMEMS HF PRESSURE MEASUREMENT SYSTEM CardioMEMSTM HF System Patient Electronics System (PES) REF CM1100 Merlin.net Patient Care Network (PCN) Heart Failure Web Application (Which interfaces with CardioMEMSTM HF System Patient Electronics System (PES) and CardioMEMSTM HF System Hospital Electronics System (HES)). Model Number MN6000

DateJan 17, 2025 (initiated Oct 5, 2024)
StatusOpen, Classified
ReasonAs a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
Root causeUnder Investigation by firm
ActionOn 11/14-18/24, the firm sent an updated Notification informing customer that Abbott is recommending replacement of the affected Patient Electronics System (PES) that are resisting correction of their configuration file. On 12/13-16/2024, the firm sent an updated Notification for the Hospital Electronic Systems (HES) regarding inaccurate Pulmonary Artery (PA) reading. On October 15, 2024, Abbott initiated a database correction to return affected PES to their intended configuration files. After continued monitoring, Abbott has identified a subset of units that is resisting correction and is re
Quantity605 units
DistributionWorldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NE, NH, NJ, NV, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and the countries of Australia, and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.