ClearanceBrief
Recall Z-0742-2025 ยท EOQ Bronchoscope (Flexible Or Rigid)

Noah Medical

Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.

DateDec 20, 2024 (initiated Oct 28, 2024)
StatusOpen, Classified
ReasonBronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.
Root causeComponent change control
ActionOn 10/28/24, recall notices were sent to customers who were told the firm would contact them to arrange inspection and replacement of the power supply system. Share this notification with any other healthcare professionals who may use the system. Complete and return the response form. Firm customer Service: customerservice@noahmed.com, 888-325-6624
Quantity18
DistributionUS: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223144Galaxy System
Noah Medical Corp., Mar 1, 2023