ClearanceBrief
Recall Z-0351-2025 ยท GCP Endoscope, Ac-Powered And Accessories

Olympus Corporation of the Americas

Brand Name: Olympus Product Name: Cystoscope Outer Sheath Model/Catalog Number: WA22810A Software Version: N/A Product Description: Outer sheaths can be combined with rigid Telescopes. Component: N/A

DateNov 8, 2024 (initiated Sep 13, 2024)
StatusOpen, Classified
ReasonOlympus is removing the statement of compatibility with a GreenLight Laser for BPH therapy from the Instruction for Use (IFU) due to the potential for a damaged tip during use of a laser probe.
Root causeUnder Investigation by firm
ActionOn October 11, 2024 URGENT: MEDICAL DEVICE CORRECTION letters were sent to consignees. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. Therefore, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Examine your inventory and identify the above-listed device. 3. You should not use the Cystoscope Outer sheath, 22.5 French with a GreenLight Laser for BPH therapy until further notice from Olympus. Add a copy of this letter to any on-hand copies of the Instructions for Use. 4. Olympu
Quantity738 units
DistributionUS Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K790071ENDOSCOPE AND ACCESSORIES
Olympus Corp., Apr 2, 1979