Recall Z-0679-2025 ยท EOQ Bronchoscope (Flexible Or Rigid)

Noah Medical

Galaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)

DateDec 13, 2024 (initiated Oct 25, 2024)
StatusOpen, Classified
ReasonNot all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.
Root causeDevice Design
ActionOn 10/25/24, recall notices were sent to customers who were asked to do the following: A copy of this letter should be kept with the instructions for use which adds additional warnings and precautions to the section Working Channel Tools/Instruments on page 23 of the Instructions for Use. CAUTION: If the bending section of the bronchoscope is articulated sharply and advancement of a biopsy tool encounters increased insertion resistance, straighten the bending section as much as possible to enable resumed advancement of the tool. Use of excessive force during instrument insertion may damage t
Quantity26
DistributionUS Nationwide Distribution: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223144Galaxy System
Noah Medical Corp., Mar 1, 2023