Recall Z-0322-2025 ยท DPS Electrocardiograph

Baxter Healthcare Corporation

Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph

DateNov 7, 2024 (initiated Oct 11, 2024)
StatusOpen, Classified
ReasonThere is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Root causeSoftware design
ActionBaxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 10/11/2024 via USPS First Class Mail. The notice explained the issue, potential hazard to the patient, and provided instructions for the to the system IT administrator to correct the issue which is already available in the current version of software.
Quantity399 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120865RSCRIBE ELECTROCARDIOGRAPH
Mortara Instrument, Inc., Jul 19, 2012