Recall Z-0327-2025 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Baxter Healthcare Corporation

Baxter AK 98 Dialysis Machine, Hemodialysis Delivery System, Product Code 955607. As of February 2025, the firm name is Vantive.

DateNov 13, 2024 (initiated Oct 16, 2024)
StatusOpen, Classified
ReasonBaxter is aware of several recalls by other manufacturers related to the potential risk of exposure to non-dioxin-like (NDL) polychlorinated biphenyl acids (PCBAs) and NDL polychlorinated biphenyls (PCBs) when using certain peritoneal dialysis and hemodialysis devices.
Root causeNonconforming Material/Component
ActionBaxter sent an IMPORTANT MEDICAL DEVICE CORRECTION notice to its consignees on 10/16/2024 via USPS. Until Baxter has transitioned to newly manufactured platinum-cured silicone tubing components in AK 98 machines, the firm recommends these actions: - If you have new AK 98 machines with a runtime less than 500 hours, consider using other available machines including AK 98 machines with a runtime greater than 500 hours, as PCB and PCBA levels decrease over time of use. - If no alternative is available, continue to use the AK 98 to ensure your patients have access to dialysis treatment. - D
Quantity1100 units
DistributionUS Nationwide distribution and worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K232467CleanCart A, CleanCart C, AK 98 Hemodialysis Machine
Baxter Healthcare Corporation, Sep 14, 2023