ClearanceBrief
Recall Z-0562-2025 ยท KDJ Set, Administration, For Peritoneal Dialysis, Disposable

Baxter Healthcare Corporation

Baxter MiniCap Extended Life PD Transfer Set with Twist Clamp, Part Number 5C4483; use in Peritoneal Dialysis

DateNov 27, 2024 (initiated Oct 23, 2024)
StatusOpen, Classified
ReasonThere has been an increase in the number of complaints related to the separation of the female connector from the main body of the transfer set.
Root causeUnder Investigation by firm
ActionBaxter issued a SAFETY ALERT notice to its consignees on 10/23/2024 via USPS first class mail. The notice explains the issue and potential hazard involved. The notice addressed to dialysis providers indicated that if a patient contacts your clinic experiencing separation of the transfer set, to replace their transfer set and contact Baxter. Notices addressed to peritoneal patients provided the following instructions: "If you experience a separation of the transfer set, please close the transfer set twist clamp and do not cut the patient line. Contact your doctor and/or nurse immediately to
Quantity62,718
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K192705MiniCap Extended Life PD Transfer Sets
Baxter Healthcare, Oct 25, 2019