ClearanceBrief
Recall Z-0408-2025 ยท OKV Obstetrical Kit

AVID Medical, Inc.

Halyard POST PARTUM TUBAL LIGATION - Medical convenience kits Model Number: LANC026-19

DateNov 14, 2024 (initiated Sep 20, 2024)
StatusOpen, Classified
ReasonSponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions
Root causeProcess control
ActionAvid Medical issued Urgent: Voluntary Medical Device Recall letter via email on 9/20/24. Letter states reason for recall, health risk and action to take: If you received any of the batches/lots identified in APPENDIX 1, please review your inventory, and discard the entire kit. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. Distributors: If any of the affected kits identified in Appendix 1 were further distributed, please ensure all end-users are appropriately notified of this produc
Quantity12 eaches
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.