Recall Z-0298-2025 ยท OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation

Boston Scientific Corporation

Boston Scientific POLARx BALLOON CATHETER LT 28MM OUS, Material Number M004CRBS2100

DateNov 12, 2024 (initiated Oct 10, 2024)
StatusOpen, Classified
ReasonBoston Scientific is updating the instructions for use of their POLARx and POLARx FIT Cryoablation Balloon Catheters related to Atrio-esophageal Fistula Risk.
Root causeDevice Design
ActionBoston Scientific disseminated a letter to Physicians/Healthcare Professionals on 10/10/2024 via courier. The letter notified its consignees of updates to the instructions for use (IFU) for the POLARx and POLARx FIT Cryoablation Catheter related to the risk of atrioesophageal (AE) fistula. The firm requested that the information be shared with clinicians who use the use the Boston Scientific POLARx Cryoablation System, including the POLARx Catheter, the POLARx FIT Catheter and the SMARTFREEZE Console, and to share the communication with any other organization to which these devices may have b
Quantity20290 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.