Recall Z-0464-2025 ยท MTF Total, Prostate Specific Antigen (Noncomplexed & Complexed) For Detection Of Prostate Cancer
Siemens Healthcare Diagnostics, Inc.
IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)
| Date | Nov 14, 2024 (initiated Sep 30, 2024) |
| Status | Open, Classified |
| Reason | The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU). |
| Root cause | Under Investigation by firm |
| Action | An Urgent Medical Device Correction letter dated 9/30/24 was sent to customers. Customer Actions " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, that may have been affected. " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution IMMULITE 2000 and IMMULITE 2000 XPi kit lots D168 and above meet the IFU high |
| Quantity | 7,332 units |
| Distribution | Worldwide - US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.