Recall Z-0318-2025 ยท LLZ System, Image Processing, Radiological

GE Healthcare

GE Healthcare Centricity Universal Viewer Zero Footprint, Model Numbers: 1) 2089507-083, GDZH-JINWAN-Centricity 2) 2089507-093, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407

DateNov 7, 2024 (initiated Oct 3, 2024)
StatusOpen, Classified
ReasonGE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate patient data.
Root causeSoftware Manufacturing/Software Deployment
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to it consignees on 10/03/2024 via letter issued through a traceable means. The notice explained the issue and provided the following actions be to be implemented until the devices can be corrected: Option 1: Restrict access to these systems from the internet. or Option 2: Implement additional safeguards such as a VPN. For questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or a local Service Representative. GE HealthCare will correct all affected products at no cost. A GE
Quantity120 units
DistributionWorldwide distribution - United States Nationwide and the countries of Algeria, Argentina, Brazil, China, Dominican Republic, France, Germany, India, Indonesia, Ireland, Israel, Japan, Kuwait, Malaysia, Mexico, Panama, Qatar, Russia, Saudi Arabia, Singapore, Spain, United Kingdom.