Recall Z-0280-2025 ยท MOQ Battery, Replacement, Rechargeable

Pro-Dex Inc

KLS Martin Battery Pack, REF: KLS BP2 001, and KLS BP2 040 (bulk case of 40), used in conjunction with the KLS Martin MaxDriver 2.0 surgical driver, REF: KLS-SD-2000

DateOct 31, 2024 (initiated Sep 23, 2024)
StatusOpen, Classified
ReasonThe packaging of a sterile battery, intended to power a surgical screwdriver, may contain Tyvek shavings inside the sterile package.
Root causeComponent design/selection
ActionOn 9/23/24, an "URGENT" recall notice was emailed to the customer informing of the product, issue, risk and actions to be taken. Customers were instructed to do the following: 1) Please immediately cease distribution and use of these lots. Inventory and distribution records should be evaluated to ensure that affected product is removed from any potential point of use. 2) Product should be returned to the recalling firm. 3) The Recall Acknowledgement and Return Form should be completed and returned via email to Recall@Pro-Dex.com or FAX 949.769.3281. Questions or concerns can be directed t
Quantity2000
DistributionUS distribution to FL only.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.