ClearanceBrief
Recall Z-0368-2025 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Stryker Corporation

Stryker Neptune SafeAir Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-046-000

DateNov 13, 2024 (initiated Oct 4, 2024)
StatusOpen, Classified
ReasonThere is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Root causeUnder Investigation by firm
ActionStryker issued Urgent Medical Device Recall Letter on 10/04/24 via Fed'X to Risk Manager, Materials Manager, OR Director. Letter states reason for recall, health risk and action to take: 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt
Quantity1,996,010 eaches
DistributionWorldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143145SafeAir Smoke Pencil
Lina Medical Aps, Dec 4, 2014